Agreement Between 10-2 and 24-2C Visual Field Test Protocols for Detecting Glaucomatous Central Visual Field Defects

Abstract

Precis

Moderate to substantial agreement between 10-2 and 24-2C perimetry for detecting central field defects suggests that adding central test points to the 24-2 protocol may improve efficiency of visual field (VF) testing for glaucoma management.

Purpose

The purpose of this study was to assess agreement between Humphrey Visual Field Analyzer 10-2 and 24-2C test protocols for detecting glaucomatous defects in the central 10 degrees of the visual field (CVFDs).

Materials and methods

VFs from 165 eyes of 18 healthy individuals, 12 glaucoma suspects and 62 glaucoma patients who completed 10-2 and 24-2C VF testing protocols within 6 months were included. CVFDs on 10-2 and 24-2C (within the central 22 points) test grids required a cluster of 3 contiguous points with P<5%, 5%, and 1% or <5%, 2%, and 2% within a hemifield on the total deviation (TD) or pattern deviation (PD) plot. Cohen kappa (k) was used to assess agreement between 10-2 and 24-2C test grids in identifying CVFDs. Specificity of each testing strategy was assessed in VFs from healthy eyes.

Results

CVFDs in suspect and glaucoma eyes were combined and reported as localized to superior, inferior or both hemifields based on TD and PD plots for 10-2 and 24-2C test grids. Moderate to substantial agreement was observed between 10-2 and 24-2C grids for detecting any CVFD from PD (k=0.551) and TD (k=0.651) plots. Specificity was high in healthy eyes ranging from 0.94 to 1.0 for both test protocols.

Conclusion

Substantial agreement for identifying CVFDs using the 24-2C and 10-2 protocols suggests that combining tests by adding central test points to the 24-2 test grid may supplant the need for 2 perimetry regimens for detecting central and peripheral glaucomatous VF damage.