A velocity map imaging spectrometer for electron?ion and ion?ion coincidence experiments with synchrotron radiation

Abstract

We have built a velocity imaging (VMI) spectrometer optimized for angle-resolved photoionization experiments with synchrotron radiation (SR) in the VUV and soft X-tay range. The spectrometer is equiped with four electrostatic lenses that focus the charged photoionization products onto a position-sensitive multi-hit delay-line anode. The use of two additional electrostatic lens elements as compared to the standard design of Eppink and Parker [T.J.B. Eppink and D.H. Parker, Rev. Sci. Instrum. 68 (1997) 3477] provides better focusing of an extended interaction region, which is crucial for most SR applications. Furthermore, the apparatus is equipped with a second micro-channel plate detector opposite to the VMI spectrometer, enabling electron-ion coincidence experiments and thereby mass-resolved ion spectroscopy independent of the time structure of the synchrotron radiation. First results for the photofragmentation of CO2 molecules are presented.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View