- Zhou, Zhiguo;
- Chen, Haixi;
- Wei, Wei;
- Zhou, Shanghui;
- Xu, Jingbo;
- Wang, Xifu;
- Wang, Qingguo;
- Zhang, Guixiang;
- Zhang, Zhuoli;
- Zheng, Linfeng
Purpose
The purpose of this study was to evaluate the image quality and radiation dose in computed tomography urography (CTU) images acquired with a low kilovoltage peak (kVp) in combination with an adaptive statistical iterative reconstruction (ASiR) algorithm.Methods
A total of 45 subjects (18 women, 27 men) who underwent CTU with kV assist software for automatic selection of the optimal kVp were included and divided into two groups (A and B) based on the kVp and image reconstruction algorithm: group A consisted of patients who underwent CTU with a 80 or 100 kVp and whose images were reconstructed with the 50% ASiR algorithm (n=32); group B consisted of patients who underwent CTU with a 120 kVp and whose images were reconstructed with the filtered back projection (FBP) algorithm (n=13). The images were separately reconstructed with volume rendering (VR) and maximum intensity projection (MIP). Finally, the image quality was evaluated using an image score, CT attenuation, image noise, the contrast-to-noise ratio (CNR) of the renal pelvis-to-abdominal visceral fat and the signal-to-noise ratio (SNR) of the renal pelvis. The radiation dose was assessed using volume CT dose index (CTDIvol), dose-length product (DLP) and effective dose (ED).Results
For groups A and B, the subjective image scores for the VR reconstruction images were 3.9±0.4 and 3.8±0.4, respectively, while those for the MIP reconstruction images were 3.8±0.4 and 3.6±0.6, respectively. No significant difference was found (p>0.05) between the two groups' image scores for either the VR or MIP reconstruction images. Additionally, the inter-reviewer image scores did not significantly differ (p>0.05). The mean attenuation of the bilateral renal pelvis in group A was significantly higher than that in group B (271.4±57.6 vs. 221.8±35.3 HU, p<0.05), whereas the image noise in group A was significantly lower than that in group B (7.9±2.1 vs. 10.5±2.3 HU, p<0.05). The CNR and SNR in group A were both significantly higher than those in group B (53.61±24.74 vs. 32.30±6.52 for CNR; 38.13±19.86 vs. 21.76±4.85 for SNR; all p<0.05). The CTDIvol, DLP and ED in group A were significantly lower than those in group B (9.26±2.77 vs. 16.19±5.60 mGy for CTDIvol; 368.86±119.38 vs. 674.38±239.37 mGy×cm-1 for DLP; 5.53±1.79 vs. 10.12±3.59 mSv for ED; all p<0.05).Conclusions
The low kVp CTU images with 50% ASiR reconstruction exhibit sufficient image quality and facilitate up to a 44% radiation dose reduction.