- Zemel, Babette;
- Wasserman, Halley;
- Kelly, Andrea;
- Fan, Bo;
- Shepherd, John;
- Lappe, Joan;
- Gilsanz, Vicente;
- Oberfield, Sharon;
- Winer, Karen;
- Kalkwarf, Heidi
BACKGROUND: Bone mineral content (BMC) and areal-bone mineral density (aBMD) measurements of the lumbar spine (LS) and whole body less head (WBLH) by dual energy X-ray absorptiometry (DXA) are recommended for bone health assessment in children. Intermachine differences were not considered previously in formulating these recommendations. METHODOLOGY: DXA measurements of the LS, WBLH, total hip, femoral neck and distal 1/3 radius from the Bone Mineral Density in Childhood Study were examined. Healthy children, ages 6 to 16 years, from five clinical centers participated. The same spine, whole body, and femur phantoms were measured on each Centers DXA machine. Percentage of individuals with low BMC or aBMD (Z-score < -1.5) was determined. Clinical center differences were evaluated by analysis of covariance adjusting for height and BMI Z-score, calcium intake, physical activity, Tanner stage and bone age. Logistic regression assessed odds of low BMC or aBMD across clinical centers. RESULTS: Significant differences among Clinical Centers (p < 0.05) were evident in adjusted mean BMC and aBMD Z-scores (n = 1503) for all skeletal sites. WBLH BMC and aBMD Z-scores had the greatest range across centers (-0.13 to 0.24, and -0.17 to 0.56, respectively). The percentage of children with Z-scores less than -1.5 varied among Clinical Centers from 1.9 [95%CI 0.8, 4.5] to 8.1 [95%CI 5.7, 11.3] for WBLH BMC, 1.1 [95%CI 0.4, 3.5] to 6.3 [95%CI 3.8, 10.1] for WBLH aBMD, and from 4.4 [95%CI 2.8, 7.0] to 12.6 [95%CI 9.3, 16.9] for distal 1/3 radius aBMD. For each skeletal site except total hip aBMD and femoral neck BMC, at least one center had significantly lower odds of low bone density. CONCLUSIONS: By design, our reference ranges capture intermachine variability. Most clinical centers dont know where their machine falls within the range of intermachine variability, and this may affect diagnosis of children evaluated for conditions that threaten bone health. Total hip scans showed the least, and whole body scans showed the most intermachine variability. Pediatric bone health assessment recommendations should recognize intermachine differences and address this important issue.