Four‐dimensional computed tomography of the left ventricle, Part I: Motion artifact reduction

Published Web Location

https://doi.org/10.1002/mp.15709
Abstract

Purpose

Standard four-dimensional computed tomography (4DCT) cardiac reconstructions typically include spiraling artifacts that depend not only on the motion of the heart but also on the gantry angle range over which the data was acquired. We seek to reduce these motion artifacts and, thereby, improve the accuracy of left ventricular wall positions in 4DCT image series.

Methods

We use a motion artifact reduction approach (ResyncCT) that is based largely on conjugate pairs of partial angle reconstruction (PAR) images. After identifying the key locations where motion artifacts exist in the uncorrected images, paired subvolumes within the PAR images are analyzed with a modified cross-correlation function in order to estimate 3D velocity and acceleration vectors at these locations. A subsequent motion compensation process (also based on PAR images) includes the creation of a dense motion field, followed by a backproject-and-warp style compensation. The algorithm was tested on a 3D printed phantom, which represents the left ventricle (LV) and on challenging clinical cases corrupted by severe artifacts.

Results

The results from our preliminary phantom test as well as from clinical cardiac scans show crisp endocardial edges and resolved double-wall artifacts. When viewed as a temporal series, the corrected images exhibit a much smoother motion of the LV endocardial boundary as compared to the uncorrected images. In addition, quantitative results from our phantom studies show that ResyncCT processing reduces endocardial surface distance errors from 0.9 ± 0.8 to 0.2 ± 0.1 mm.

Conclusions

The ResyncCT algorithm was shown to be effective in reducing motion artifacts and restoring accurate wall positions. Some perspectives on the use of conjugate-PAR images and on techniques for CT motion artifact reduction more generally are also given.