Understanding the mechanical behavior of intrauterine devices during simulated removal
Published Web Location
https://doi.org/10.1016/j.contraception.2024.110399Abstract
Objective
To evaluate differences based on intrauterine device (IUD) frame geometry in force, and stress, and strain at the stem/arms junction during simulated IUD removal.Study design
We manufactured injection-molded frame models for three Nova-T IUDs (Mirena [model M]; Liletta [model L]; Kyleena [model K]) and a Tatum-T IUD (Paragard [model P]) at two-times scaling. We created a custom fixture to simulate the uterus and used a screw-driven machine to pull models at various displacement rates through the 10 cm fixture cavity to measure force and strain and calculate stress at the IUD stem/arms junction. We tested models at 30 mm/min and higher displacement rates for exploratory analyses. We used Mann-Whitney U test for statistical testing.Results
We completed testing at 30 mm/min using five of each Nova-T model and nine model P samples. Resistance against the cavity walls created significantly more force on model P (11.83, interquartile range [IQR] 11.61-12.31) than any Nova-T model samples (p < 0.001). The smaller model K created slightly more median stress (MPa) than the larger model M (0.36 [IQR 0.33-0.38] and 0.79 [IQR 0.76-0.80], respectively, p = 0.008); model P samples generated significantly more median stress than other models (1.70 [IQR 1.67-1.77], p < 0.001). Strain plots demonstrated permanent deformation for some samples during IUD removal simulation. We tested 20 samples at various higher displacement rates up to 2500 mm/min, with stress notably increasing for model P samples with increasing rates. No fractures occurred.Conclusions
Force and stress at the stem/arms junction are higher with Tatum-T-shaped compared to Nova-T-shaped IUD models under the same testing conditions, and a higher speed of extraction causes more stress.Implications
Sharp corners create vulnerability under static and fatigue loading in structural components due to increased local stresses. Our findings suggest that IUDs with Tatum-T frames should be removed slowly to minimize the stress at the stem/arms junction. Future studies can provide more information if performed with commercially available products.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
For improved accessibility of PDF content, download the file to your device.
Enter the password to open this PDF file:
File name:
-
File size:
-
Title:
-
Author:
-
Subject:
-
Keywords:
-
Creation Date:
-
Modification Date:
-
Creator:
-
PDF Producer:
-
PDF Version:
-
Page Count:
-
Page Size:
-
Fast Web View:
-
Preparing document for printing…
0%