Multipole Shimming of Permanent Magnets Using Harmonic Corrector Rings

Abstract

Shimming systems are required to provide sufficient field homogeneity for high resolution NMR. In certain specialized applications, such as rotating-field NMR and portable (ex-situ) NMR, permanent magnet-based shimming systems can provide considerable advantages. We present a simple two-dimensional shimming method based on harmonic corrector rings which can provide arbitrary multipole order shimming corrections. Results demonstrate, for example, that quadrupolar order shimming improves the linewidth by up to and order of magnitude. An additional order of magnitude reduction is in principle achievable by ultilizing this shimming method for z-gradient correction and higher order xy gradients.