Frequency domain diffuse optical spectroscopy with a near-infrared tunable vertical cavity surface emitting laser.
Published Web Location
https://doi.org/10.1364/OE.26.021033Abstract
We present an approach for performing frequency domain diffuse optical spectroscopy (fd-DOS) utilizing a near-infrared tunable vertical cavity surface emitting laser (VCSEL) that enables high spectral resolution optical sensing in a miniature format. The tunable VCSEL, designed specifically for deep tissue imaging and sensing, utilizes an electrothermally tunable microelectromechanical systems topside mirror to tune the laser cavity resonance. At room temperature, the laser is tunable across 14nm from 769 to 782nm with single mode CW output and a peak output power of 1.3mW. We show that the tunable VCSEL is suitable for use in fd-DOS by measuring the optical properties of a tissue-simulating phantom over the tunable range. Optical properties were recovered within 0.0006mm-1 (absorption) and 0.09mm-1 (reduced scattering) compared to a broadband fd-DOS reference system. Our results indicate that tunable VCSELs may be an attractive choice to enable high spectral resolution optical sensing in a wearable format.