Development of a Soluble KIT Electrochemical Aptasensor for Cancer Theranostics

Abstract

An electrochemical sensor based on a conformation-changing aptamer is reported to detect soluble KIT, a cancer biomarker, in human serum. The sensor was fabricated with a ferrocene-labeled aptamer (Kd < 5 nM) conjugated to a gold electrode. Quantitative KIT detection was achieved using electrochemical impedance spectroscopy (EIS) and square-wave voltammetry (SWV). EIS was used to optimize experimental parameters such as the aptamer-to-spacer ratio, aptamer immobilization time, pH, and KIT incubation time, and the sensor surface was characterized using voltammetry. The assay specificity was demonstrated using interfering species and exhibited high specificity toward the target protein. The aptasensor showed a wide dynamic range, 10 pg/mL-100 ng/mL in buffer, with a 1.15 pg/mL limit of detection. The sensor also has a linear response to KIT spiked in human serum and successfully detected KIT in cancer-cell-conditioned media. The proposed aptasensor has applications as a continuous or intermittent approach for cancer therapy monitoring and diagnostics (theranostics).

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.