Long-term assessment of the safety and effectivity of the mini-jupette sling: 5-year follow-up of the original series

Abstract

Background

In 2017, a prospective multicenter, multinational, investigational pilot study was conducted examining outcomes using a novel surgical technique, the Mini-Jupette sling, for the management of erectile dysfunction (ED) patients with climacturia and/or minimal stress urinary incontinence (SUI) after prostate procedures. Climacturia has been reported in up to 64% of patients following radical prostatectomy (RP). We sought to report the 5-year outcomes from this original cohort to assess long-term safety and effectivity of the mini-jupette sling in the treatment of ED and concomitant mild SUI and/or climacturia.

Methods

This is a single-arm, multicenter, retrospective, observational study. We identified patients who were enrolled in the previous multicenter study with post-RP ED and climacturia and/or mild SUI- 2 PADS PER DAY (PPD) and underwent inflatable penile prosthesis (IPP) insertion with simultaneous placement of a mini-jupette sling. Data were collected including current PPD, subjective improvement in climacturia/SUI, complications, need for revision of IPP or additional urinary incontinence surgery, and date of most recent follow-up. SPSS was used for statistical analysis.

Results

Of the original 38 patients, 5 have since died and 10 were lost to follow-up with 23/38 (61%) patients available for evaluation of long-term outcomes. The average follow-up time was 59 months (SD =8.8) with a mean age of 69 years (SD =6.8). Most patients (n=21, 91%) had subjective improvement of SUI and climacturia. One patient with persistent bothersome incontinence underwent artificial urinary sphincter (AUS) placement in 2018 with no complications, while the other is still considering a repeat procedure due to minor but persistent SUI. The mean PPD decreased from 1.4 preoperatively to 0.4 at a mean of 5 years of follow-up. Most patients reported satisfaction in their urinary symptoms with 91% and 73% reporting improvement in SUI and climacturia respectively, compared to 86% and 93% respectively in the original series. One (4.3%) patient had an IPP revision for pump malfunction. There were no device infections reported.

Conclusions

The mini-jupette sling appears to be a safe and effective procedure with durable improvements in SUI and climacturia at 5 years of follow-up.