Improving Technical Outcome in Osteochondral Allograft Transplantation of the Femoral Trochlea: The Role of Contoured Guides, Experience Level, Graft Size, and Graft Position

Abstract

Background

The use of coring instrumentation for osteochondral allograft (OCA) transplantation of the femoral trochlea is challenging due to the complex topography of this anatomical area.

Purpose

We sought to determine the effect of flat guides versus guides contoured to the surface of the trochlea on graft step-off in trochlear OCAs performed on a foam bone model. We also determined the effect of surgeon experience level and graft size on step-off.

Study design

Controlled laboratory study.

Methods

Allograft harvesters were produced in 3 different sequential sizes with either a flat or a contoured undersurface. The guides matched one another in all aspects of shape and size except for the undersurface contour. The contoured undersurface generally matched the surface of the bone model trochlea but was not customized to that surface. A total of 72 foam femora were obtained. Identical trochlear stellate lesions of 3 different sizes (small, medium, and large) were created using 3-dimensional printed surface templates. A total of 6 surgeons (3 attending and 3 resident surgeons) performed OCAs of each trochlear lesion. Each surgeon performed 1 graft for each size and each guide type (n = 6 per surgeon). A specialized digital caliper was prepared that allowed the measurement of graft step-off to within 0.01 mm at a distance of 5 mm between the 2 sensors. The measurements were performed at 8 positions around the clockface of the grafts.

Results

Grafts performed with the contoured guides had a mean step-off of 0.43 ± 0.37 mm. Grafts performed with the flat guides had a mean step-off of 0.74 ± 0.78 mm (P < .0001). Experience level did not have a significant effect on graft step-off (P = .81). There was no identifiable pattern indicating higher step-off at any one position on the clockface.

Conclusion

In this study, contoured grafts had significantly lower step-offs compared with flat guides. Experience level, clockface position, and graft size did not affect step-off.