Reliability investigation of the instrument transfer function calibration technique based on binary pseudo-random array standards

Abstract

The reliability of the instrument transfer function (ITF) calibration technique based on binary pseudo-random array (BPRA) standards is investigated and demonstrated in application to interferometric microscopes. We demonstrate the linearity of the calibration (that is, independence of the ITF calibration on the standards root-mean-square roughness) via comparison of the ITF measurements with a number of artifacts with the etched depth varying from 30 nm to 120 nm. We also show that the calibration does not depend on the surface reflectivity, at least in the range between ∼36% and ∼80%. The criteria for selection of the geometrical parameters of the BPRA standard design appropriate for a particular interferometric microscope arrangement (including optical magnification), as well as the data acquisition and analysis procedures for different applications are also discussed.