Skip to main content
eScholarship
Open Access Publications from the University of California

UC Irvine

UC Irvine Previously Published Works bannerUC Irvine

3D-printed dissolvable inserts for efficient and customizable fabrication of NMR transceiver coils

Abstract

We describe a simplified method for improving the reproducibility of transceiver coil fabrication for nuclear magnetic resonance (NMR) through single-use templates made from 3D-printed polymer forms. The utility of dissolvable inserts for achieving performance enhanced resonators (DIAPERs) is tested herein by a comparison of RF homogeneity along the rotor axis for variable-pitch solenoids with different inter-turn spacing. Simulated B1 field profiles are compared to experimental homogeneity measurements, demonstrating the potential of this approach for making NMR coils quickly and reproducibly.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View