Assessment of trabecular bone structure using MDCT: comparison of 64- and 320-slice CT using HR-pQCT as the reference standard
Skip to main content
eScholarship
Open Access Publications from the University of California

Assessment of trabecular bone structure using MDCT: comparison of 64- and 320-slice CT using HR-pQCT as the reference standard

  • Author(s): Issever, Ahi S.
  • Link, Thomas M.
  • Kentenich, Marie
  • Rogalla, Patrik
  • Burghardt, Andrew J.
  • Kazakia, Galateia J.
  • Majumdar, Sharmila
  • Diederichs, Gerd
  • et al.
Abstract

The aim of our study was to perform trabecular bone structure analysis with images from 64- and 320-slice multidetector computed tomography (MDCT) and to compare these with high-resolution peripheral computed tomography (HR-pQCT). Twenty human cadaver distal forearm specimens were imaged on a 64- and 320-slice MDCT system at 120 kVp, 200 mA and 135 kVp, 400 mA (in-plane pixel size 234 µm; slice thickness 500 µm). HR-pQCT imaging was performed at an isotropic voxel size of 41 µm. Bone volume fraction (BV/TV), trabecular number (Tb.N), thickness (Tb.Th) and separation (Tb.Sp) were computed. MDCT-derived BV/TV and Tb.Sp were highly correlated (r = 0.92–0.96, p < 0.0001) with the corresponding HR-pQCT parameters. Tb.Th was the only structure measure that did not yield any significant correlation. The 64- and 320-slice MDCT systems both perform equally well in depicting trabecular bone architecture. However, because of constrained resolutions accurate derivation of trabecular bone measures is limited to only a subset of microarchitectural parameters.