Quantitative measurement of atheroma burden: reproducibility in serial studies of atherosclerotic femoral arteries

Abstract

Objectives

This study aims to evaluate the reproducibility of measures of plaque morphology in serially acquired black-blood MRI of untreated atherosclerotic femoral arteries.

Materials and methods

MR studies was obtained from 42 timepoints, on 12 patients with known femoral artery atherosclerosis. Images with a 3D isotropic FLASH with DANTE-prepared black blood contrast (DASH) at a 3-T scanner were acquired at baseline, within 1 week, and at 1 month. Six of the patients were scanned additionally at 6 months. Inter-scan and inter-observer variations of arterial area/volume measurements were evaluated.

Results

Measurement of vessel area, lumen area, wall area and wall volume showed inter-scan intraclass correlation coefficients (ICC) ranging from 0.92 to 0.97 for 3 scans, 0.91-0.97 for 4 scans, and inter-observer ICCs of 0.89-0.96. Among 3 scans, the coefficients of variance (CV) for the vessel area, lumen area, wall area and wall volume were 4.1%, 6.5%, 7.5%, and 4.4%. CVs among 4 scans ranged from 4.4% to 7.9%, and interobserver CVs ranged from 6.1% to 11.8% for the different area/volume measurements.

Conclusion

DASH MRI is useful for quantifying atherosclerotic vessel area and volume of femoral arteries with low variability among serial repeated scans and between observers.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Main Content
For improved accessibility of PDF content, download the file to your device.