Comparison of Predicted and Experimentally Measured Aerosol Deposition Efficiency in BALB/C Mice in a New Nose-Only Exposure System
Abstract
The use of respiratory tract dosimetry predictions to estimate the desired aerosol exposure concentration required for a specific target deposited dose in animal inhalation studies has been hindered by: (1) a lack of species/strain specific quantitative respiratory tract anatomy; and (2) verification by comparison of calculated and experimentally measured deposited doses. Using recent literature on tracheobronchial and pulmonary respiratory tract anatomy, dosimetry predictions for the Balb/c mouse were compared with deposited doses for 0.5, 1.0, and 2.0 micrometer diameter aerosols in a newly available nose-only exposure system. Spatial and temporal aerosol port to port uniformity of this nose-only exposure system was within ± 10% of the mean. Dosimetry predictions were in agreement with the measured mean deposited doses for the aerosols tested. © 2009 Taylor & Francis Group, LLC.