Quantification of myocardial blood flow using non–ECG‐triggered MR imaging
Published Web Location
https://doi.org/10.1002/mrm.25451Abstract
Purpose
MR myocardial perfusion imaging is dependent on reliable electrocardiogram (ECG) triggering for accurate measurement of myocardial blood flow (MBF). A non-ECG-triggered method for quantitative first-pass imaging may improve clinical feasibility in patients with poor ECG signal. The purpose of this study is to evaluate the feasibility of a non-ECG-triggered method for myocardial perfusion imaging in a single slice.Methods
The proposed non-ECG-triggered technique uses a saturation-recovery magnetization preparation and golden-angle radial acquisition for integrated arterial input function (AIF) measurement. Image based self-gating with a temporal resolution of 42.6 ms is used to generate a first-pass image series with consistent cardiac phase. The AIF is measured using beat-by-beat T1 estimation of the ventricular blood pool. The proposed technique was performed on 14 healthy volunteers and compared against a conventional ECG-triggered dual-bolus acquisition.Results
The proposed method produced MBF with no significant difference compared with ECG-triggered technique (mean of 0.63 ± 0.22 mL/min/g to 0.73 ± 0.21 mL/min/g).Conclusion
We have developed a non-ECG-triggered perfusion imaging method with T1 based measurement of the AIF in a single slice. In this preliminary study, our results demonstrate that MBF measured using the proposed method is comparable to the conventional ECG-triggered method.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.