Characterizing Breast Lesions Using Quantitative Parametric 3D Subharmonic Imaging: A Multicenter Study

Abstract

Rationale and objectives

Breast cancer is the leading type of cancer among women. Visualization and characterization of breast lesions based on vascularity kinetics was evaluated using three-dimensional (3D) contrast-enhanced ultrasound imaging in a clinical study.

Materials and methods

Breast lesions (n = 219) were imaged using power Doppler imaging (PDI), 3D contrast-enhanced harmonic imaging (HI), and 3D contrast-enhanced subharmonic imaging (SHI) with a modified Logiq 9 ultrasound scanner using a 4D10L transducer. Quantitative metrics of vascularity derived from 3D parametric volumes (based on contrast perfusion; PER and area under the curve; AUC) were generated by off-line processing of contrast wash-in and wash-out. Diagnostic accuracy of these quantitative vascular parameters was assessed with biopsy results as the reference standard.

Results

Vascularity was observed with PDI in 93 lesions (69 benign and 24 malignant), 3D HI in 8 lesions (5 benign and 3 malignant), and 3D SHI in 83 lesions (58 benign and 25 malignant). Diagnostic accuracy for vascular heterogeneity, PER, and AUC ranged from 0.52 to 0.75, while the best logistical regression model (vascular heterogeneity ratio, central PER, and central AUC) reached 0.90.

Conclusion

3D SHI successfully detects contrast agent flow in breast lesions and characterization of these lesions based on quantitative measures of vascular heterogeneity and 3D parametric volumes is promising.