Background
Cardiac resynchronization therapy (CRT) has been shown to improve both the functional status and mortality of heart failure patients with left bundle branch block. Multiple recent studies suggest several mechanisms for proarrhythmia associated with CRT device.Case summary
A 51-year-old male with symptomatic non-ischemic cardiomyopathy and no previous history of ventricular arrhythmias underwent placement of a biventricular cardioverter-defibrillator. The patient developed sustained monomorphic ventricular tachycardia (VT) soon after implantation. The VT recurred despite reprogramming to right ventricular only pacing. The electrical storm resolved only after a subsequent discharge from the defibrillator caused inadvertent dislodgement of the coronary sinus lead. No recurrent VT occurred throughout 10-years follow up after urgent coronary sinus lead revision.Discussion
We describe the first reported case of mechanically induced electrical storm due to the physical presence of the CS lead in a patient with a new CRT-D device. It is important to recognize mechanical proarrhythmia as a potential mechanism of electrical storm, as it may be intractable to device reprogramming. Urgent coronary sinus lead revision should be considered. Further studies on this mechanism of proarrhythmia are needed.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
For improved accessibility of PDF content, download the file to your device.
Enter the password to open this PDF file:
File name:
-
File size:
-
Title:
-
Author:
-
Subject:
-
Keywords:
-
Creation Date:
-
Modification Date:
-
Creator:
-
PDF Producer:
-
PDF Version:
-
Page Count:
-
Page Size:
-
Fast Web View:
-
Preparing document for printing…
0%