Skip to main content
eScholarship
Open Access Publications from the University of California

Preclinical assessment of rodent jumping power with a novel electrical stimulation-assisted device.

Abstract

Sarcopenia is a progressive loss of muscle mass and strength that is associated with increasing the risk of falls, musculoskeletal diseases, and chronic metabolic diseases. However, the animal models adopted to study sarcopenia face limitations since the functional tests conducted on human cannot be directly adapted to animals because the animals do not follow instructions. Moreover, current preclinical research tools for muscle function assessment, such as the rotarod, grip strength, and treadmill, have limitations, including low-intensity simple movements, evaluator subjectivity, and limited power indicators. Hence, in this study, we present a new jumping-power assessment tool in a preclinical rodent model to demonstrate muscle functions. To overcome the light weight and command issues in the rodent model, we developed an electrical stimulation-assisted jump power assessment device. Precisely, the device utilizes a load cell with a 0.1 g resolution and a 50 points/s data acquisition rate to capture the short period of the mouse jump. Additionally, interdigitated electrodes are used to electrically stimulate the mice and make them jump. While our primary focus in this article is the validation of the newly developed jump power assessment device, it is worth noting that this tool has several potential utilities. These include the phenotypic comparison of sarcopenia models, the exploration of muscle function reduction mechanisms, muscle function-related blood biomarkers, and the evaluation of drug intervention effects.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View