- Main
Miniaturized Wirelessly Powered and Controlled Implants for Multisite Stimulation.
Published Web Location
https://doi.org/10.1109/tmtt.2022.3233368Abstract
This paper presents a miniaturized implant with a diameter of only 14 mm, which houses a novel System on Chip (SoC) enabling two voltage level stimulation of up to 16 implants using a single Tx coil. Each implant can operate at a distance of 80 mm in the air through the inductive resonant link. The SoC consumes only 27 μW static power and enables two channels with stimulation amplitudes of 1.8 V and 3.3 V and timing resolution of 100 μs. The SoC is implemented in the standard 180 nm complementary metal oxide semiconductor (CMOS) technology and has an area of 0.75 mm × 1.6 mm. The SoC comprises an RF rectifier, low drop-out regulator (LDO), error detection block, clock data recovery, finite state machine (FSM), and output stage. Each implant has a PCB-defined passcode, which enables the individual addressability of the implants for synchronized therapies. The implantable device weighs only 80 mg and sizes 20.1 mm3. Tolerance of up to 70° to angular misalignment was measured at a distance of 50 mm. The efficacy of bilateral stimulation was further verified by implanting two devices on two sides of a pigs neck and performing bilateral vagus nerve stimulation (VNS), while monitoring the heart rate.
Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
Enter the password to open this PDF file:
-
-
-
-
-
-
-
-
-
-
-
-
-
-