Skip to main content
eScholarship
Open Access Publications from the University of California

UCSF

UC San Francisco Previously Published Works bannerUCSF

A Millimeter-scale Single Charged Particle Dosimeter for Cancer Radiotherapy.

Abstract

This paper presents a millimeter-scale CMOS 64×64 single charged particle radiation detector system for external beam cancer radiotherapy. A 1×1 μm2 diode measures energy deposition by a single charged particle in the depletion region, and the array design provides a large detection area of 512×512 μm2. Instead of sensing the voltage drop caused by radiation, the proposed system measures the pulse width, i.e., the time it takes for the voltage to return to its baseline. This obviates the need for using power-hungry and large analog-to-digital converters. A prototype ASIC is fabricated in TSMC 65 nm LP CMOS process and consumes the average static power of 0.535 mW under 1.2 V analog and digital power supply. The functionality of the whole system is successfully verified in a clinical 67.5 MeV proton beam setting. To our' knowledge, this is the first work to demonstrate single charged particle detection for implantable in-vivo dosimetry.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View