- Nakazato, Ryo;
- Slomka, Piotr J;
- Fish, Mathews;
- Schwartz, Ronald G;
- Hayes, Sean W;
- Thomson, Louise EJ;
- Friedman, John D;
- Lemley, Mark;
- Mackin, Maria L;
- Peterson, Benjamin;
- Schwartz, Arielle M;
- Doran, Jesse A;
- Germano, Guido;
- Berman, Daniel S
Background
Obesity is a common source of artifact on conventional SPECT myocardial perfusion imaging (MPI). We evaluated image quality and diagnostic performance of high-efficiency (HE) cadmium-zinc-telluride parallel-hole SPECT MPI for coronary artery disease (CAD) in obese patients.Methods and results
118 consecutive obese patients at three centers (BMI 43.6 ± 8.9 kg·m(-2), range 35-79.7 kg·m(-2)) had upright/supine HE-SPECT and invasive coronary angiography > 6 months (n = 67) or low likelihood of CAD (n = 51). Stress quantitative total perfusion deficit (TPD) for upright (U-TPD), supine (S-TPD), and combined acquisitions (C-TPD) was assessed. Image quality (IQ; 5 = excellent; < 3 nondiagnostic) was compared among BMI 35-39.9 (n = 58), 40-44.9 (n = 24) and ≥45 (n = 36) groups. ROC curve area for CAD detection (≥50% stenosis) for U-TPD, S-TPD, and C-TPD were 0.80, 0.80, and 0.87, respectively. Sensitivity/specificity was 82%/57% for U-TPD, 74%/71% for S-TPD, and 80%/82% for C-TPD. C-TPD had highest specificity (P = .02). C-TPD normalcy rate was higher than U-TPD (88% vs 75%, P = .02). Mean IQ was similar among BMI 35-39.9, 40-44.9 and ≥45 groups [4.6 vs 4.4 vs 4.5, respectively (P = .6)]. No patient had a nondiagnostic stress scan.Conclusions
In obese patients, HE-SPECT MPI with dedicated parallel-hole collimation demonstrated high image quality, normalcy rate, and diagnostic accuracy for CAD by quantitative analysis of combined upright/supine acquisitions.