- Szurman, Peter;
- Mansouri, Kaweh;
- Dick, H Burkhard;
- Mermoud, Andre;
- Hoffmann, Esther M;
- Mackert, Marc;
- Weinreb, Robert N;
- Rao, Harsha Laxmana;
- Seuthe, Anna-Maria;
- Haus, Arno;
- Reifschneider, Eugen;
- Rickmann, Annekatrin;
- Wakili, Philip;
- Emmerich, Caroline;
- Hoogmartens, Caroline;
- Lambert, Kristina;
- Januschowski, Kai;
- Elling, Matthias;
- Schultz, Tim;
- Kersten-Gomez, Inga;
- Lorenz, Katrin;
- GIers, Bert;
- Beck, Anna;
- Nobl, Matthias;
- Freissinger, Sigrid
Aim
To investigate the safety and performance of a telemetric suprachoroidal intraocular pressure (IOP) sensor (EYEMATE-SC) and the accuracy of its IOP measurements in open angle glaucoma (OAG) patients undergoing simultaneous non-penetrating glaucoma surgery (NPGS).Methods
Prospective, multicentre, open-label, single-arm, interventional clinical trial. Twenty-four eyes of 24 patients with OAG regularly scheduled for NPGS (canaloplasty or deep sclerectomy) were simultaneously implanted with an EYEMATE-SC sensor. Six-month follow-up on the sensor's safety and performance as well as on the level of agreement between the EYEMATE-SC measurements and IOP measurements with Goldmann applanation tonometry (GAT).Results
The eyes underwent canaloplasty (n=15) or deep sclerectomy (n=9) and achieved successful implantation of the sensor. No device migration, dislocation or serious device-related complications occurred. A total of 367 comparisons were included in the IOP agreement analysis. The overall mean difference between GAT and EYEMATE-SC measurements was 1.31 mm Hg (lower limit of agreement (LoA) 7.55 mm Hg; upper LoA -4.92 mm Hg). The maximum difference of 2.5 mm Hg ±3.96 (LoA 0.30-2.29) was reached on day 10 and continuously improved to an agreement of -0.15 mm Hg ±2.28 (LoA -1.24 to 0.89) after 6 months. Accordingly, the percentage of eyes within an IOP difference of ±5 mm Hg improved from 78% (day 3) to 100% (6 months).Conclusions
After 6 months, the EYEMATE-SC sensor was safe and well tolerated, and allowed continual IOP monitoring.Trial registration number
NCT03756662.