BACKGROUND: Intraocular pressure (IOP) monitoring in glaucoma management is evolving with novel devices. We investigated the reproducibility of 24 hour profiles on two consecutive days and after 30 days of self-measurements via telemetric IOP monitoring. METHODS: Seven primary patients with open-angle glaucoma previously implanted with a telemetric IOP sensor in one eye underwent automatic measurements throughout 24 hours on two consecutive days (day 1 and day 2). Patients wore an antenna adjacent to the study eye connected to a reader device to record IOP every 5 min. Also, self-measurements in six of seven patients were collected for a period of 30 days. Analysis included calculation of hourly averages to correlate time-pairs of day 1 versus day 2 and the self-measurements vers day 2. RESULTS: The number of IOP measurements per patient ranged between 151 and 268 on day 1, 175 and 268 on day 2 and 19 and 1236 during 30 days of self-measurements. IOP time-pairs of automatic measurements on day 1 and day 2 were significantly correlated at the group level (R=0.83, p<0.001) and in four individual patients (1, 2, 6 and 7). IOP time-pairs of self-measurements and day 2 were significantly correlated at the group level (R=0.4, p<0.001) and in four individual patients (2, 5, 6 and 7). CONCLUSIONS: Twenty-four hour automatic measurements of IOP are correlated on consecutive days and, though to a lesser degree, with self-measurements. Therefore a virtual 24-hour IOP curve might be constructed from self-measurements. Both options provide an alternative to frequent in-office IOP measurements.