Skip to main content
eScholarship
Open Access Publications from the University of California

UC Irvine

UC Irvine Previously Published Works bannerUC Irvine

Alarming increase in HbA1c and near misdiagnosis of diabetes mellitus resulting from a clinical laboratory instrument upgrade and haemoglobin variant

Published Web Location

https://casereports.bmj.com/content/2018/bcr-2018-225358.full
No data is associated with this publication.
Abstract

A 29-year-old woman was referred for new-onset diabetes mellitus after her glycosylated haemoglobin (HbA1c) was found to be 10.2%. Three years earlier, the patient's HbA1c-measured by the same clinical laboratory-had been 5.5%. The newer HbA1c was discordant with fasting glucose levels and a lack of diabetes-associated symptoms. The laboratory reported that their assay methodology remained unchanged and also that no haemoglobin variants were detected. Further investigation, however, revealed, first, that the patient carried a haemoglobin alpha chain mutation (Hb Wayne) that can sometimes cause assay interference and, second, that although the laboratory's assay methodology had not changed, their assay instrument had. Depending on assay methodology, haemoglobin variants can cause HbA1c assay interference and the presence of these variants may not be detected by the performing laboratory. Interference may not only be dependent on assay methodology but also on the assay instrument used.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Item not freely available? Link broken?
Report a problem accessing this item