Macular Ganglion Cell/Inner Plexiform Layer Measurements by Spectral Domain Optical Coherence Tomography for Detection of Early Glaucoma and Comparison to Retinal Nerve Fiber Layer Measurements

Abstract

Purpose

To evaluate the performance of ganglion cell layer/inner plexiform layer (GCL/IPL) measurements with spectral-domain optical coherence tomography (Cirrus HD-OCT) for detection of early glaucoma and to compare results to retinal nerve fiber layer (RNFL) measurements.

Design

Cross-sectional prospective diagnostic study.

Methods

We enrolled 99 subjects, including 59 eyes with glaucoma (47 subjects) (mean deviation >-6.0 dB) and 91 normal eyes (52 subjects). Patients underwent biometry and peripapillary and macular OCT imaging. Performance of the GCL/IPL and RNFL algorithms was evaluated with area under receiver operating characteristic curves (AUC), likelihood ratios, and sensitivities/specificities adjusting for covariates. Combination of best parameters was explored.

Results

Average (SD) mean deviation in the glaucoma group was -2.5 (1.9) dB. On multivariate analyses, age (P < 0.001) and axial length (P = 0.03) predicted GC/IPL measurements in normal subjects. No significant correlation was found between average or regional GC/IPL thickness and respective outer retina (OR) thickness measurements (P > 0.05). Average RNFL thickness performed better than average GCL/IPL measurements for detection of glaucoma (AUC = 0.964 vs 0.937; P = 0.04). The best regional measures from each algorithm (inferior quadrant RNFL vs minimum GCL/IPL) had comparable performances (P = 0.78). Entering the GC/IPL to OR ratio into prediction models did not enhance the performance of the GCL/IPL measures. Combining the best parameters from each algorithm improved detection of glaucoma (P = 0.04).

Conclusions

Regional GCL/IPL measures derived from Cirrus HD-OCT performed as well as regional RNFL outcomes for detection of early glaucoma. Using the GC/IPL to OR ratio did not enhance the performance of GCL/IPL parameters. Combining the best measures from the 2 algorithms improved detection of glaucoma.