A liquid scintillation detector for radioassay of gadolinium-loaded liquid scintillator for the LZ Outer Detector

Published Web Location

https://www.sciencedirect.com/science/article/pii/S0168900219307119?via%3Dihub
No data is associated with this publication.
Abstract

We report on the design and performance of the LUX-ZEPLIN (LZ) “Screener”, a small liquid scintillator detector consisting of ≈23 kg of LAB-based gadolinium-loaded liquid scintillator (GdLS) to be used in the LZ Outer Detector. The Outer Detector will be filled with 17.3 tonnes of GdLS and will surround the central liquid xenon time projection chamber of LZ. Its primary function will be to tag neutron events in the liquid xenon which could mimic a WIMP dark matter signal. To meet the deadtime requirements for the Outer Detector, the radioimpurity levels in the GdLS must be kept below ≲0.07 mBq/kg. This background level corresponds to a rate of ≈50Hz above an energy threshold of 100 keV. The Screener was operated in the ultra-low-background environment of the former LUX water shield in the Davis Laboratory at the Sanford Underground Research Facility for radioassay of the GdLS. Careful selection of detector materials and use of ultra-low-background PMTs allows the measurement of a variety of radioimpurities. The 14C/12C ratio in the scintillator is measured to be (2.83±0.06(stat.)±0.01(sys.))×10−17. Use of pulse shape discrimination allows the concentration of isotopes throughout the 238U, 235U, and 232Th chains to be measured by fitting the collected spectra from α and β events. We find that equilibrium is broken in the 238U and 232Th chains and that a significant portion of the contamination in the GdLS results from decays in the 227Ac subchain of the 235U series. Predictions for the singles rate in the Outer Detector are presented. The rate from radioimpurities above 100 keV in the GdLS is estimated to be 97.9±6.4Hz, with 65.5±1.9Hz resulting from α-decays.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Item not freely available? Link broken?
Report a problem accessing this item