Evaluating the performance of the Roche FEN2 fentanyl immunoassay and its clinical implementation: The role of LDT-based mass spectrometry testing

Abstract

Introduction

While laboratory-developed tests (LDTs) using liquid chromatography tandem mass spectrometry (LC-MS/MS) are widely employed to support the development of FDA-cleared drug immunoassays, their significance in the clinical implementation and evaluation of such assays is often overlooked. This paper reports on the important role of LC-MS/MS LDTs in demonstrating improved performance of the Roche FEN2 fentanyl immunoassay compared with the Thermo DRI fentanyl immunoassay.

Methods

The FEN2 assay was implemented according to the manufacturer's instructions and its performance was compared to the existing DRI assay using LC-MS/MS as a reference. Clinical sensitivity and specificity were determined using 250 consecutive random patient specimens. Spiking experiments were conducted to determine cross-reactivity with 31 fentanyl analogs. Select DRI false-positive samples were analyzed by the FEN2 assay via time-of-flight mass spectrometry method (LC-QTOF).

Results

The FEN2 assay showed improved clinical sensitivity compared to the DRI (98% vs 61%) in 250 consecutive patient samples due to its ability to detect norfentanyl. It also showed better clinical specificity by correctly classifying select DRI false-positive results. Upon implementation in clinical practice, the FEN2 resulted in a higher screening positivity rate than the DRI (17.3% vs 13.3%) and a greater LC-MS/MS confirmation rate of immunoassay-positive samples (96.8% vs 88.8%, respectively).

Conclusion

The use of LC-MS/MS LDTs demonstrated that the FEN2 assay has greater clinical sensitivity and is less prone to false-positives than the DRI assay. These findings support the use of FEN2 in routine clinical practice and emphasize the role of mass spectrometry-based LDTs in clinical toxicology testing.