Comparison of six serum analyte workflows using routine liquid chromatography-tandem mass spectrometry methods at multiple laboratories with the Cobas i 601 analyzer, an automated mass spectrometry system
Skip to main content
eScholarship
Open Access Publications from the University of California

UC San Diego

UC San Diego Previously Published Works bannerUC San Diego

Comparison of six serum analyte workflows using routine liquid chromatography-tandem mass spectrometry methods at multiple laboratories with the Cobas i 601 analyzer, an automated mass spectrometry system

Abstract

OBJECTIVES: Liquid chromatography-tandem mass spectrometry (LC-MS/MS) offers substantial analytical advantages over traditional immunoassay methods, including greater specificity and robustness. Conventional LC-MS/MS methods are labor-intensive, requiring highly trained, dedicated laboratory staff. The Cobas® i 601 analyzer (Roche Diagnostics International Ltd., Rotkreuz, Switzerland) is an automated, random-access, mass spectrometry (MS) system designed to improve workflows and reduce turnaround times. This study evaluated turnaround times for analyses with the i 601 analyzer vs. routine LC-MS/MS using batch processing. METHODS: We performed a workflow analysis for six sets of analytes (mycophenolic acid, antibiotics, antiepileptics, voriconazole, testosterone, and 25-hydroxyvitamin D) in seven laboratories across Europe, North America, and Asia. We recorded batch size, result turnaround times, and hands-on times (including quality control) for 93 batches analyzed on routine LC-MS/MS. These were compared to equivalent measurements generated using the i 601 analyzer in batch mode as part of a prototype pilot study at Ludwig Maximilian University, Munich. RESULTS: Turnaround times were shorter with the i 601 analyzer vs. routine LC-MS/MS for all tested workflows, independent of batch size and analyte workflow. Median time from batch start to final result availability with the i 601 analyzer was 6.8-fold shorter vs. routine LC-MS/MS (1 h, 34 min vs. 10 h, 37 min, respectively). In addition, median total hands-on time was 19-fold shorter with the i 601 analyzer vs. routine LC-MS/MS (5 vs. 93 min). CONCLUSIONS: Compared with routine LC-MS/MS methods, the i 601 analyzer substantially reduced turnaround and hands-on times for common analytes, regardless of batch size.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.