An implantable 700μW 64-channel neuromodulation IC for simultaneous recording and stimulation with rapid artifact recovery

  • Author(s): Johnson, BC
  • Gambini, S
  • Izyumin, I
  • Moin, A
  • Zhou, A
  • Alexandrov, G
  • Santacruz, SR
  • Rabaey, JM
  • Carmena, JM
  • Muller, R
  • et al.
Abstract

© 2017 JSAP. We present an 180nm HV CMOS IC for concurrent neural stimulation and recording that combines 64 low-noise recording front-ends and 4 independent stimulators multiplexed to any of the 64 channels. The stimulators have 5mA peak current, 12V compliance and dynamic power management to maximize efficiency. Co-design of the stimulation and recording subsystems resulted in 100mV of recording linear range, 70nV/rtHz noise, and a rapid 1ms (single-sample) artifact recovery during stimulation.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.