Skip to main content
eScholarship
Open Access Publications from the University of California

UC Davis

UC Davis Previously Published Works bannerUC Davis

Prospective Cross-Sectional Study of Repeatability of Peripapillary Capillary Density Measurement Using Optical Coherence Tomography Angiography in Eyes With Optic Nerve and Retinal Vascular Pathology

Abstract

Background

Optical coherence tomography angiography (OCTA) is a new noninvasive imaging modality that provides high resolution images of the optic nerve head and peripapillary retinal capillary vasculature which can be affected by optic nerve or retinal pathologies. High repeatability of peripapillary capillary density measurement using OCTA has been demonstrated in normal eyes and eyes with glaucoma. The purpose of our study was to quantify the repeatability of peripapillary capillary density measurement using OCTA in both normal eyes and eyes with optic atrophy, optic disc edema, and retinal vasculopathy.

Methods

This prospective cross-sectional study enrolled 31 patients (59 eyes) including 16 eyes with optic nerve pathology (7 with disc edema from papilledema and 9 with optic atrophy), 35 eyes with retinal vascular disease, and 8 normal eyes. All eyes were imaged twice (30 minutes apart) with the Optovue AngioVue OCTA instrument to obtain 4.5 × 4.5 mm peripapillary scans. Scans were considered good quality if signal strength was 6 or greater. The OCTA parameters obtained include the radial peripapillary capillary (RPC) density of the whole disc, inside the disc, peripapillary region, and the 4 quadrants of the disc (superior, nasal, inferior, and temporal). A Student's t test was used to compare means. Intraclass correlation coefficient (ICC) was calculated to measure repeatability.

Results

Repeatability of RPC density measurements for all regions analyzed demonstrated good to excellent repeatability for the whole cohort {ICC for the whole image was 0.915 (95% confidence interval [CI] = 0.855-0.951)}; ICC for the peripapillary region was 0.945 (95% CI = 0.905-0.969). In the subset of eyes with good image quality (i.e., signal strength ≥ 6), ICC was slightly higher for all regions, with excellent repeatability of the peripapillary region (ICC was 0.971 [95% CI = 0.943-0.986]). Conversely, for eyes with poor image quality scans (i.e., signal strength < 6), ICC was lower, corresponding to moderate to good repeatability for most parameters. For the subset of eyes with optic atrophy, disc edema from papilledema or retinal vasculopathy, all had good to excellent repeatability of the vessel density of the entire disc (ICC values were 0.954 [95% CI = 0.804-0.990], 0.921 [95% CI = 0.711-0.982], and 0.895 [95% CI = 0.788-0.951, respectively]) and of the peripapillary region (ICC values were 0.980 [95% CI = 0.904-0.996], 0.966 [95% CI = 0.854-0.993], and 0.916 [95% CI = 0.827-0.961], respectively).

Conclusions

The peripapillary capillary density measurement obtained using a commercial OCTA instrument is highly repeatable in eyes with optic nerve atrophy, disc edema from papilledema, or retinal vasculopathy.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View