Clinical Experience with Cone-Beam CT Navigation for Tumor Ablation
Published Web Location
https://doi.org/10.1016/j.jvir.2014.10.049Abstract
Purpose
To describe clinical use and potential benefits of cone-beam computed tomography (CT) navigation to perform image-guided percutaneous tumor ablation.Materials and methods
All ablations performed between February 2011 and February 2013 using cone-beam CT navigation were included. There were 16 patients who underwent 20 ablations for 29 lesions. Cone-beam CT ablation planning capabilities include multimodality image fusion and tumor segmentation for visualization, depiction of the predicted ablation zones for intraprocedural planning, and segmentation of the ablated area for immediate verification after treatment. Number and purpose of cone-beam CT scans were examined. The initial ablation plan, defined as number of probes and duration of energy delivery, was recorded for the 20 of the 29 lesions ablated. Technical success and local recurrences were recorded. Primary and secondary effectiveness rates were calculated.Results
Image fusion was used for 16 lesions, and intraprocedural ultrasound was used for 4 lesions. Of the 20 ablations, where the ablation plans were recorded, there was no deviation from the plan in 14 ablations. In the remaining 6 ablations, iterative planning was needed for complete tumor coverage. An average of 8.7 cone-beam CT scans ± 3.2 were performed per procedure, including 1.3 ± 0.5 for tumor segmentation and planning, 1.7 ± 0.7 for probe position confirmation, and 3.9 ± 2 to ensure complete coverage. Mean follow-up time was 18.6 months ± 6.5. Ablations for 28 of 29 lesions were technically successful (96.5%). Of ablations performed with curative intent, technical effectiveness at 1 month was 25 of 26 lesions (96.1%) and 22 of 26 lesions (84.6%) at last follow-up. Local tumor progression was observed in 11.5% (3 of 26 lesions).Conclusions
Cone-beam CT navigation may add information to assist and improve ablation guidance and monitoring.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
For improved accessibility of PDF content, download the file to your device.
Enter the password to open this PDF file:
File name:
-
File size:
-
Title:
-
Author:
-
Subject:
-
Keywords:
-
Creation Date:
-
Modification Date:
-
Creator:
-
PDF Producer:
-
PDF Version:
-
Page Count:
-
Page Size:
-
Fast Web View:
-
Preparing document for printing…
0%